@iandobbie

Latest updated from photmetrics.

Just a little update. We have been able to duplicate the same behaviour on our cameras as well.
The camera is reporting exposure time in nanoseconds. At 100ms of exposure for the sensitivity
port (this is observed only for the sensitivity port), the camera reports 99.997938ms. The PVCam
doesn’t round it up but shows only value before the decimal point.

Engineering is looking into it. Unfortunately, the fix will take a while. We will update you again once
we know something more.

I suggest until such a time as it is fixed we just add 1ms to all exposure times to prevent missing triggers.