Medtronic MyCareLink Smart 25000 all versions are...
High severity
Unreviewed
Published
May 24, 2022
to the GitHub Advisory Database
•
Updated May 22, 2025
Description
Published by the National Vulnerability Database
Dec 14, 2020
Published to the GitHub Advisory Database
May 24, 2022
Last updated
May 22, 2025
Medtronic MyCareLink Smart 25000 all versions are vulnerable to a race condition in the MCL Smart Patient Reader software update system, which allows unsigned firmware to be uploaded and executed on the Patient Reader. If exploited an attacker could remotely execute code on the MCL Smart Patient Reader device, leading to control of the device.
References